Flow control for droplet generation using syringe pumps and pressure-based flow controllers 

Micrometer droplets and particles are widely used in a broad range of industries. Droplet production using microfluidic systems was implemented for applications where monodispersity is of high importance. Within micrometer-sized channels, one droplet at a time is generated, allowing for the production of monodisperse droplets. With this level of control, applications that would not have been possible before, have also emerged, such as digital PCR and single cell encapsulation within droplets.

The importance of flow control for droplet production 

Syringe pumps are commonly used for generating droplets in microfluidic experiments, but can show limited flow control. As a result, the droplet size, which is proportional to the flow rate, is affected. Therefore, flow control for droplet generation is critical for having repeatable reactor volume and reproducible results. An alternative to syringe pumps is pressure-based flow controllers. These show that high-precision flow control, high reaction time, and flow monitoring is possible. 

Syringe pumps are commonly used for generating droplets in microfluidic experiments, but can show limited flow control. As a result, the droplet size, which is proportional to the flow rate, is affected. Therefore, flow control for droplet generation is critical for having repeatable reactor volume and reproducible results. An alternative to syringe pumps is pressure-based flow controllers. These show that high-precision flow control, high reaction time, and flow monitoring is possible. 

Flow control for droplet
Figure 1: Generation of droplets using a) batch method and b) microfluidic method 
Complete analysis comparing syringe pumps and pressure-based flow controllers
Analysis between Microfluidic syringe pumps and flow EZ

We compare the production of water-in-oil emulsions using microfluidic syringe pumps and Fluigent pressure-based flow controllers. Droplet size stability and the time required to reach several droplet diameters are determined for the two instruments. 

Droplet production under identical experimental conditions 

Flow control for droplet
Figure 2: Scheme of droplet generation using the Flow EZ and syringe pumps 

Flow rate stability over time when generating droplets 

To determine the droplet stability over time achieved with the different instruments, droplets are generated within a microfluidic chip, working with constant water and oil flow rates of 2.0 µL/min and 1.5 µL/min, respectively. 

Figure 3 shows the flow rate of water and oil as a function of time for standard syringe pumps and for the Flow-EZ, the pressure-based flow controller. The average flow rates, standard deviations, and coefficients of variation were also calculated. The water and HFE flow rate variations using the Flow EZ are lower than when using standard syringe, confirming the enhanced flow stability over time of pressure-based flow controllers. 

Droplet stability over time
Figure 3: Flow rate stability of water and HFE over time with the Flow EZ and syringe pump 

Table 1: Water and HFE average flow rate, STD and CV with the Flow EZ and syringe pump 

droplet-generatio
Figure 4: Droplet generation in the microfluidic chip channels 

To determine the effect of flow control for droplet generation, and thus the effect of flow rate on droplet stability, a picture of the channels was taken every 10 seconds, and the average diameter over 5 minutes was calculated for each device. To better compare between the different instruments, we use normalized values. 

The figure at left shows the normalized diameter variation over time. We can observe that droplets are more stable over time using the Flow EZ compared to the standard syringe pump. This aligns with the results obtained for flow rate stability and confirms that the more precise the flow rate, the more monodisperse the droplets will be. 

Normalized droplet diameter variation over time
Figure 5: Normalized droplet diameter as a function of time using the Flow EZ and syringe pump 

Impact of precise flow control on droplet diameter over time 

It can also be of interest to be able to quickly move from one droplet size to another. We can determine the response time for reaching a specific droplet diameter. The water flow rate is varied, while the oil flow rate is kept constant at 1.5 µL/min. The set flow rates were the following: 1 µL/min – 4 µL/min – 1 µL/min – 2 µL/min. 

Figure 6 shows the results obtained using the Flow EZ. We observe that the set flow rates are reached in a few seconds. After reaching a set flow rate, we can observe the droplet size that results.  

In the analysis of the effect of flow control on droplet generation, droplets of 89 µm, 101 µm and approximately 110 µm in diameter were obtained via flow rates of 1 µL/min, 2 µL/min and 4 µL/min, respectively, when using the Flow EZ. 

Droplet diameter change over time - Flow control for droplet
Figure 6: Droplet diameter change as a function of time using the Flow EZ. The water flow rates are varied over time, while the HFE flow rate was kept constant 
droplet production with syringe pump vs pressure flow controller

Figure 7: Time required to change from one water flow rate to another using the Flow EZ and syringe pumps. 

Figure 7 shows the water flow rate as a function of time for the syringe pump and the Flow EZ, and the table below shows the time required to move from one flow rate to the other after starting the system. We observe that the maximum time required to switch to another flow rate using the Flow EZ and the syringe pump are 6 s and 93 s, respectively. 

We can therefore see that the Flow EZ is the instrument with the shortest response time. This allows users to rapidly change from one droplet size to another. 

FEZ 345 mbar and Syringe pump table

Conclusion

In analyzing the effect of flow control on droplet generation, more stable flow rates were observed when generating droplets with the Flow EZ, resulting in more monodisperse droplets compared to syringe pumps. These results confirm the relationship between flow rate and droplet size.  

The difference in response time between the different devices was also determined. Better response times were observed when using the Flow EZ, allowing for responsive control over droplet size during an experiment.  

Micrometer-size droplets and particles are widely used in a broad range of industries. Fluigent provides complete, cost-effective solutions for the production of monodisperse droplets., allowing users to control droplet size and frequency by adjusting flow parameters. 

Related content

  • Droplet and particle generation in microfluidics

    Microfluidic Droplet Production Method

    Read more
  • Application notes

    Encapsulation of multiple emulsions in a single droplet 

    Read more
  • White Papers

    Droplet-based Microfluidics – White Paper

    Read more
  • White Papers

    Double emulsion for the generation of microcapsules – a Review​

    Read more
  • Support & Tools

    Droplet Size Calculator

    Read more

Related products

  • Fluigent PX Read

For more information or a technical discussion

Contact us
Logo fluigent green and blue

67 avenue de Fontainebleau
94 270 Le Kremlin-Bicêtre

Research

  • Instruments
  • Microfluidic Research Applications
  • Software solutions
  • Packages

Industrial

  • Industrie produkte

Resources

  • Application notes
  • Customer case studies
  • Expertise reviews
  • Interviews & Testimonials
  • Webinars

Support & tools

  • Documentation
  • Download software

Company

  • Über uns
  • Team
  • Events
  • Fluigent Newsletter
  • Careers

Legal

  • Terms & Conditions of Sale
  • Legal Terms & Privacy Policy